Canty Solidsizer -
: Measuring proppants, frac sand, and ores.
| Task | Action | |------|--------| | Change lens | Power off camera, unscrew lens, align new lens, refocus | | Adjust gap | Flow on → loosen lock screw → turn micrometer → tighten lock screw | | Clean windows | Isolate cell → flush warm DI water 10 min → dry with nitrogen | | Run daily check | Capture reference beads (50 µm) – D50 should be 48–52 µm | | Update software | Backup \CantyData\ folder first → run installer as admin | canty solidsizer